22
MODEL CLARITY II
SECTION 5.0
PROGRAMMING THE ANALYZER
5.3.3. Procedure: Configure Outputs.
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
1.
Press MENU. The main menu screen appears. Choose
Program
.
2.
Choose
Outputs
.
3.
Choose
Output Configure
.
4.
Choose
Output1
or
Output2
. This screen appears only in instruments
having dual input.
5.
Choose
Sensor1
or
Sensor2
. Either sensor can be assigned to either
output.
6.
Make the appropriate settings:
a. Choose
4-20 mA
or
0-20 mA
.
b.
Choose
Yes
or
No
for output dampening.
7.
The display returns to the screen in step 3. Select the other output or
press EXIT to return to the previous screen. To return to the main display,
press MENU followed by EXIT.
5.3.4. Procedure: Assigning Values to the Low and High Current Outputs (Output Ranging)
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
1.
Press MENU. The main menu screen appears. Choose
Program
.
2.
Choose
Outputs
.
3.
Choose
Output Range
.
4.
Choose
Output1
or
Output2
. This screen appears only in instruments
having dual input.
5.
Make the appropriate settings.
a.
Assign a value to the low current (
0 mA
or
4 mA
) output.
b.
Assign a value to the high current (
20 mA
) output.
6.
The display returns to the screen in step 4. Select the other output or
press EXIT to return to the previous screen. To return to the main display,
press MENU followed by EXIT.
Calibrate
Hold
P
Pr
ro
og
gr
ra
am
m
Display
Output Config?
O
Ou
ut
tp
pu
ut
t1
1
Output2
OutM is for?
S
Se
en
ns
so
or
r1
1
Sensor2
Output Range
O
Ou
ut
tp
pu
ut
t
C
Co
on
nf
fi
ig
gu
ur
re
e
O
Ou
ut
tp
pu
ut
ts
s
Alarms
Measurement
>>
Calibrate
Hold
P
Pr
ro
og
gr
ra
am
m
Display
O
Ou
ut
tp
pu
ut
t
R
Ra
an
ng
ge
e
Output Configure
O
Ou
ut
tp
pu
ut
ts
s
Alarms
Measurement
>>
Output Range
O
Ou
ut
tp
pu
ut
t1
1
Output2