24
MODEL CLARITY II
SECTION 5.0
PROGRAMMING THE ANALYZER
5.4.3 Procedure: Configuring Alarms
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
1.
Press MENU. The main menu screen appears. Choose
Program
.
2.
Choose
Alarms
.
3.
Choose
Alarm Configure
.
4.
To disable all alarms, choose
Y
(yes). The display returns to the screen
shown in step 2. If you want to use the alarm relays, choose
N
. Go to step 5.
5.
Choose Alarm 1 (
AL1
), Alarm 2 (
AL2
), or Alarm 3 (
AL3
).
6.
For
AL1
or
AL2
a.
Choose
Sensor 1
or
Sensor 2
. For a single input configuration, the
Sensor 1 Sensor 2
screen does not appear.
b.
Choose
High
or
Low
.
c.
Set the alarm
Deadband
.
7.
The display returns to the
Alarm Configure?
screen. Select another
alarm or press EXIT to return to the previous screen. To return to the main
display, press MENU followed by EXIT.
8.
For
AL3
a.
Choose
Sensor1
,
Sensor2
, or
Fault
.
b.
For
Sensor1
, select
High
or
Low
and set the deadband.
c.
Choosing
Fault
means
AL3
will activate when a sensor or analyzer
fault exists. There is no user setting to make.
9.
The display returns to the
Alarm Configure?
screen. Select another
alarm or press EXIT to return to the previous screen. To return to the main
display, press MENU followed by EXIT.
Calibrate
Hold
P
Pr
ro
og
gr
ra
am
m
Display
Alarm Setpoints
A
Al
la
ar
rm
m
C
Co
on
nf
fi
ig
gu
ur
re
e
Alarm Config?
A
AL
L1
1
AL2
AL3
AL1 is for?
S
Se
en
ns
so
or
r1
1
Sensor2
Outputs
A
Al
la
ar
rm
ms
s
Measurement
>>
AL3 is for?
F
Fa
au
ul
lt
t
Sensor1
Sensor2
Disable all
Alarms?
Y
N
N