44
MODEL CLARITY II
SECTION 6.0
CALIBRATION
6.7.2 Procedure
1.
First, calibrate the sensor. See Section Section 6.2, 6.3, or 6.4.
C
Ca
al
li
ib
br
ra
at
te
e
Hold
Program
Display
Calibrate?
S
Se
en
n1
1
Sen2
Output
2.
Press the MENU key. The main menu appears. Choose
Calibrate
.
3.
If the analyzer has dual input choose
Sen1
(sensor 1) or
Sen2
(sensor 2).
If the analyzer has single input, choose
Sensor
.
4.
Choose
Enter TSS Data
.
5. The display prompts the user to enter TSS for point 1 (
Pt1
). The units
shown in the second line will be the units selected in Section 5.5.3. Press
ENTER.
6. The display prompts the user to enter the turbidity for point 1. Press
ENTER.
7. The display prompts the user to enter TSS for point 2 (
Pt2
). Press
ENTER.
8. The display prompts the user to enter the turbidity for point 2. Press
ENTER.
9.
The screen at left appears if the calibration was successful. Press
ENTER.
10. If the calibration was unsuccessful, the screen at left appears. Repeat
steps 6 through 9, checking for data entry errors.
11. If the intercept on the NTU axis is negative, the analyzer will display the
low turbidity limit. See Section 6.6.1 for more information.
12. To return to the main display, press MENU then EXIT.
SN TSS Pt1?
X
X
XXppm
SN Pt2?
XXXppm
Turbid?
X
X
X.XNTU
SN Pt1?
XXXppm
Turbid?
X
X
X.XNTU
SN TSS Pt2?
X
X
XXppm
Calibrate?
S
Se
en
ns
so
or
r
Output
Calibrate
Enter TSS Data
>>
SN Calculation
Complete
Calibration
Error
Low Turbidity
Limit
x.xxxNTU