2.4 INSTALLATION — SENSOR
Unscrew the nut on the side of the debubbler. Insert the sensor in the mouth of the measuring chamber. Be sure
the pin on the debubbler lines up with the hole in the sensor. Replace the nut. Remove the protective cap from
the sensor and screw the cable onto the receptacle. The plug and receptacle are keyed for proper alignment.
2.5 SAMPLE POINT
Locate the sample tap to minimize pickup of sediment or air. See Figure 2-6. If possible, install a sampling port
that extends one or two inches (25 - 50 mm) into the pipe. Use ¼ inch OD rigid plastic tubing. Avoid soft plastic
tubing if possible. To reduce sample lag time, install the debubbler and flow chamber as close to the sample tap
as possible.
FIGURE 2-6. Sampling for Turbidity