Emerson Process Management GmbH & Co. OHG
2-15
2
Installation
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
07/2017
Fig. 2-10:
Allocation of terminals
1 Terminals for signal cables
2 Power EMI filter
3 4 cable entries for power and signal cables
4
Power terminals with integrated fuses
5 Ethernet service port and USB connectors
6 Ethernet network connection
2.5 Installation – Electrical
Unused entries during installation have to be
provided with plugs.
Installation under CSA certification requires
the use of approved conduits and suitable
metric-to-NPT adaptors. .
Both plugs and cable entries must be installed
properly and fixed by applying the prescribed
torque.
The instrument provides internal screw termi-
nals for connecting power and signal cables.
This requires opening the instrument during
installation:
•
Unsrew the 20 screws located on the
enclosure flange.
•
Flap down the cover part to gain access
to the analyzer inside.
Inside the enclosure, keep all cables as short
as possible to prevent from interferences.
ATEX approved analyzers provide 4 cable
entries (M20 x 1.5) and are shipped with a
set of 2 plugs.
Owners of such analyzers are responsible to
provide suitable cable glands according all
applicable standards (e. g. IEC/EN 60079-14)
5 1 6 2 3 4