Emerson Process Management GmbH & Co. OHG
S-11
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
07/2017
S
Safety Instructions
Safety Instructions
EXPLOSION AND ELECTRICAL SHOCK HAZARD
All cables (power and signal) must end (be connected) in either a safe (non-
hazardous) area or in a protecting enclosure (e.g. Ex e junction box)!
The power and signal cables must be separated by a distance of minimum
1 cm (0.4 in) inside and outside the analyzer!
EXPLOSION HAZARD
Startup, operation and service must not be performed, before reading and
understanding all instructions!
Especially all warnings in this and the associated manuals have to be
considered! Inspection, maintenance and service must be carried out
considering all related standards e.g. for “Inspection and maintenance of
electrical installations in hazardous areas” or “Equipment repair, overhaul
and reclamation“.
EXPLOSION HAZARD: VAPOR RECOVERY APPLICATION
Consider the waiting time statement on the front door label before opening!
Disregarding the waiting time statement may result in explosion!
Temperatures of components for this application installed into the analyzer
exceed the analyzer’s temperature classification for hazardous areas! This
option is only possible for -20°C to 50°C ambient temperature and for non-
flammable sample gas.