Maintenance
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
07/2017
Modifications made on the electrical appara
-
tus affecting the integrity of the type of protec
-
tion or the temperature of the apparatus shall
be permitted only if the modified apparatus is
resubmitted to a testing station.
4.1
Verifications and Tests
4.1 Verifications and Tests
4.1.1 Routine Tests
The following tests shall be performed on
a regular basis:
EXPLOSION HAZARD BY DAMAGED ANALYZER
Do not continue to operate a damaged analyzer!
Call for service!
Violation may cause an explosion and/or personal injury or death!
4.1.1.1 Visual Inspection
The instrument shall be visually checked for
damages to the enclosure, flange and exter
-
nal analyzer components.
Especially take care that there are no scratch-
es on the flanges surfaces!
See section 4.1.1.2 for more information on
how to verify the flanges’ compliances with
the certification requirements, if the results
of the visual inspection is not clear.
Contents
4.1 Verifications and Tests
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4.2 Replacement of Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4.3 Vapor Recovery Application – Special Information. . . . . . . . . . . . . . . . . . . 4-9
4.4 Perform a calibration
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
In the case of repairs affecting the type of
protection, the parts which have been re
-
paired should be subjected to new routine
verifications and tests. These tests need not
necessarily be made by the manufacturer.
Compare the installed components with the list
shown in section 4.2 of this manual: Unauthor-
ized replacement of parts is not permitted, may
affect the safety of the analyzer and violates
the certification!