4.1 Verifications and Tests
4.1.1.2 Detailed Inspection
•
Check cable inlets (glands or conduits):
If the cable moves, e. g. thighten the cap
nut. If tightening is ineffective, replace the
cable gland or conduit.
Also consider the original manufacturer’s
instructions for maintenance!
•
Damages to the flanges cannot be ac
-
cepted, as they act as flame paths. If need
be, submit the unit to one or more of the
following measurements to confirm com
-
pliance with the certification requirements:
•
Evenness of flange surfaces:
max. accepted
0.04 mm per flange
•
Roughness Ra of flanges:
max. accepted 6.3 µm
If one or both of above measurements
show non-compliances,
•
take the analyzer out of service by dis
-
connecting all electrical connections
•
submit the unit to the Emerson service
center.