ELECTRICAL SHOCK HAZARD WHEN OPERATED OPENED
Do not operate without covers secure. Do not open while energized.
Installation requires access to live parts which can cause death or serious
injury.
For safety and proper performace this instrument must be connected to a
properly grounded three-wire source of power.
Violation may cause explosion and personnel injury!
EXPLOSION HAZARD WHEN OPEN
Do NOT operate the instrument with doors or covers open! This is permitted
only when no hazardous atmosphere is present! Depending on the local
regulation, this may require a competent hot work supervisor to issue a
hot work permit.
Use ALL 20 screws with the respective torque to fix the cover!
Violation may cause an explosion hazard!
EXPLOSION HAZARD BY HIGH PRESSURE
Risk of internal overpressure under leakage conditions!
For normal operation up to 1100 hPa (1.45 psig) sample gas pressure the
maximum flow into the containment system is 1.5 lpm for flammable gases.
For high pressure option up to 1500 hPa (7.25 psig) the maximum flow into
the enclosure is 1.0 lpm. The sum of gas flows into the enclosure including
purge gas flows not not exceed 2 l/min. Make sure to close gas supplies
before you open the front cover.
Take care of special conditions for safe use, and gas parameter specifications!