Emerson Process Management GmbH & Co. OHG
4-15
4
Maintenance
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
07/2017
Zero C
alibration..
Span C
alibration..
Advanced C
alibration..
Apply Gas..
Lock M
enus!
Acknowledgements
Pump 1:
Off
Pump 2:
Off
Ranges..
4.4 Manual Calibration
When finished, press
left
several times to
return to SELECT COMPONENT
(multi
channel analyzer only), to perform a span
calibration for another channel,
or
press
home
to return to the MEASUREMENT
SCREEN
, to finish with manual calibration
procedures.
4.4.2 Manual Span Calibration
Multi-channel unit:
Select the channel to be calibrated in SE-
LECT COMPONENT.
Before selecting any further line
make sure the required calibra-
tion gas is applied and flowing!
SPAN CALIBRATION
offers the similiar op
-
tions as ZERO CALIBRATION.
Select
“
Span Calibration..”
Span C
alibration
Cancel!
Start!
Span Gas
0.000 ppm
Concentration
0.000 ppm
Flow
0.00 l/min
Status..
Results..
Restore..