1.3.6 Power voltage limits on direct output used together with external SURGE CDN
When voltages higher than 250V between “High” and “Low” of the direct output of the IMU4000 occur a
gating will be activated.
The gating can influence the EUT current as the IMU4000 Surge circuit will be turned ON/OFF and act as a
load.
1.3.7 Voltage interruption and Variation with internal Variac EXT-TRA3000 D-V
Voltage range
0 to 260 V when EUT
power input voltage is
230V
EUT
Power
Depending on the EUT
power voltage
Frequency range without variac
DC up to 60 Hz
external Source
Frequency range with variac involved 48 Hz to 60 Hz
external Source
Nominal current
16A without internal
Variac involved
Interruption with internal variac and
linear load
maximum 12 A
maximum 16 A
< 5s
< 300 ms
Inrush current
500 A Peak
- 0%,
+30%
Interruption time
50 µs to 30 s
phase angle selectable
Amplitude of the interruptions
continuously selectable
from 0 to 100 %
IEC: 0 %, 40 %, 70 %, 80%
Phase angle for turn ON and OFF of
the EUT. Selectable in range
0 to 360°
± 5°
Voltage variation with the internal
variac
0 to 110 % maximum.
5A
± 20%
2 s to 30000 s
Voltage variation with external variac 0 to 110 % maximum.
16 A
± 20%
2 s to 30000 s
Less than 1 period
More than one period
d.c. interruption
Interruption within one
period.
Interruption longer than
one period.
Input as phase angle
Input in ms
Input in ms
Ramps
-Voltage
-Synchronisation angle
-Interruption time
Interruption for all kind of loads
UT= voltage at EUT Power 1
DIP
100 %
% UT
0 %
0 to 16 A
For interruptions of 0 to 100% and 100% to 0% the internal Variac is not involved,
therefore the test can be carried out up to 16 A. For interruption with UT =EUT Power 1
voltage not zero, the internal variac limits the EUT power current. The maximum
allowed current values are listed in the table on the next page. Please be aware that
different types of loads influence the maximum current capability.
Summary of Contents for IMU4000
Page 8: ...IMU4000 8 104...
Page 19: ...IMU4000 19 104 Derating curve for EXT IMU4000 F5...
Page 56: ...IMU4000 56 104...
Page 66: ...IMU4000 66 104...
Page 70: ...IMU4000 70 104...
Page 74: ...IMU4000 74 104...
Page 76: ...TRANSIENT 2000 76 104...
Page 78: ...TRANSIENT 2000 78 104...
Page 92: ...Appendix and Corrections 92 104 13 1 6 VARIATION Specification IEC 61000 4 11 Ed 2...
Page 94: ...Appendix and Corrections 94 104...
Page 100: ...Index 100 104...
Page 102: ...Index 102 104...
Page 104: ...Index 104 104...