E-IMU4000-E-Manual
33/104
3 Mechanical structure
3.1 General
The IMU4000 is ideal for running tests in development/test laboratory environments and for outdoor service
on larger systems. For outdoor service, the IMU4000 can be fitted into a military case.
For better understanding, the IMU4000 will be divided into two parts:
The left hand part of the IMU4000 contain the control and measurements. The left hand side of the front
panel, is called the control panel.
The right hand part contains all high voltage circuits, such us high voltage source, high voltage switches,
the impulse-forming network and the coupling / de-coupling network. This part is called the operation
panel.
Fig.3.1
The power connections of the IMU4000 and the EUT are located on the rear panel. With the EUT power
inputs on the rear side and the outputs on the front side an optimum de-coupling is guaranteed. This
arrangement allows test set-up without parallel-running cables.
The IMU4000 is available with different options:
Standard with handles on both side
as showed in Figure 3.1. This version is recommended for use in
development and EMC test laboratories.
19“ insert version
. The handles are removed and angle brackets are fixed on both sides for fitting the
IMU4000 in a 19“ rack. When the IMU4000 is equipped with an EFT circuit the EFT output must be
maximum 50 cm above the reference ground plane.
Standard with handle in a military case
. This version is recommended for outdoor EMC testing.
Summary of Contents for IMU4000
Page 8: ...IMU4000 8 104...
Page 19: ...IMU4000 19 104 Derating curve for EXT IMU4000 F5...
Page 56: ...IMU4000 56 104...
Page 66: ...IMU4000 66 104...
Page 70: ...IMU4000 70 104...
Page 74: ...IMU4000 74 104...
Page 76: ...TRANSIENT 2000 76 104...
Page 78: ...TRANSIENT 2000 78 104...
Page 92: ...Appendix and Corrections 92 104 13 1 6 VARIATION Specification IEC 61000 4 11 Ed 2...
Page 94: ...Appendix and Corrections 94 104...
Page 100: ...Index 100 104...
Page 102: ...Index 102 104...
Page 104: ...Index 104 104...