IMU4000
36/104
fig. 3.4
Fig. 3.4
At DIPS to 0 % of the power line voltage, two operating conditions can be differentiated:
a) Switch S1 is opened, the voltage of the power decreases at the EUT with the discharge constant of the
EUT (High Z at 0% = ON)
b) Some µs after switch S1 has opened, switch S2 will be closed and the EUT will be discharged via the
circuit EUT Power 2 (High Z at 0% = OFF).
AT High - Z Mode = OFF and large capacitive loads, the large capacitance will be discharged via the
internal variac at the beginning of the interruption. A large current will result, if an interruption to 0% of the
power line voltage is generated. To avoid reducing the life span of the carbon contact electrode of the
variac, it is recommended to make a short circuit with an external bridge between L2 and N of EUT Power 2.
Vrms between EUT Power 1 and EUT Power 2 must be lower than 250V. Use for EUT Power 1
and 2 equal phase L.
The maximum voltage on the inputs of EUT power 1 or EUT power 2 must be lower than 280V.
High voltages will destroy the varistors on the inputs.
Phase
EUT Power 1
EUT Power 2
EUT Power
Output
S1
S1
S2
S2
U-power
U-power
S1= ON
S2 = OFF
S1= OFF
S2= ON
S1= ON
S2= OFF
U - Power line
L
L
N
N
EUT
External or internal Variac (or
PS3)
Neutral
EUT Power 1
EUT Power 2
Summary of Contents for IMU4000
Page 8: ...IMU4000 8 104...
Page 19: ...IMU4000 19 104 Derating curve for EXT IMU4000 F5...
Page 56: ...IMU4000 56 104...
Page 66: ...IMU4000 66 104...
Page 70: ...IMU4000 70 104...
Page 74: ...IMU4000 74 104...
Page 76: ...TRANSIENT 2000 76 104...
Page 78: ...TRANSIENT 2000 78 104...
Page 92: ...Appendix and Corrections 92 104 13 1 6 VARIATION Specification IEC 61000 4 11 Ed 2...
Page 94: ...Appendix and Corrections 94 104...
Page 100: ...Index 100 104...
Page 102: ...Index 102 104...
Page 104: ...Index 104 104...