CDN2000-06-25
coupling decoupling network
8/54
i
Primary
Protection
Secondary
Protection
Household
Equipment
e.g. PC
ANSI / IEEE; IEC 61000-4-12
CWG versus Ring wave
Lightning
Figure: 1.0.2-1
The ring wave appears at the terminals of equipment (equipment ports) as a consequence of switching in
power and control lines, as well as a consequence of lightning. The single event type and the decaying
oscillatory waveform are the most significant parameters of this test.
Ring Wave Generator
Figure: 1.0.2-2
Legend:
RL
Charging resistor
Cs Impulse capacitor
S
High voltage switch
MS
Measuring equipment
Lp
Inductance of the generator
Rs
Serial resistor
RM
Shunt
Lv
Inductance of the connection
R1-3 Current limiting resistors
TM
Dividers
Cp Parallel capacitor
Test generator
S
Summary of Contents for CDN2000-06-25
Page 18: ...CDN2000 06 25 coupling decoupling network 18 54...
Page 22: ...CDN2000 06 25 coupling decoupling network 22 54...
Page 24: ...CDN2000 06 25 coupling decoupling network 24 54...
Page 30: ...CDN2000 06 25 coupling decoupling network 30 54...
Page 36: ...CDN2000 06 25 coupling decoupling network 36 54...
Page 38: ...CDN2000 06 25 coupling decoupling network 38 54...
Page 40: ...CDN2000 06 25 coupling decoupling network 40 54...
Page 46: ...CDN2000 06 25 coupling decoupling network 46 54 V Verification by EMC PARTNER 38...
Page 47: ...CDN2000 06 25 coupling decoupling network 47 54...
Page 49: ...CDN2000 06 25 coupling decoupling network 49 54 13 3 Assembly Instruction to MC Connectors...
Page 50: ...CDN2000 06 25 coupling decoupling network 50 54...
Page 52: ...CDN2000 06 25 coupling decoupling network 52 54...
Page 53: ...CDN2000 06 25 coupling decoupling network 53 54...
Page 54: ...CDN2000 06 25 coupling decoupling network 54 54...