CDN2000-06-25
coupling decoupling network
6/54
Generators to carry out combination wave tests
Figure: 1.0.3-1
Legend:
RL
Charging resistor
Cs
Impulse capacitor
S
High voltage switch
RE1 Discharge resistor
Ls
Inductance of the generator
Rs
Serial resistor
RM
Shunt
RE2 Discharge resistor
TM1 Divider
Lv
Inductance of the connection
MS
Measurement equipment
TM2 Divider
A dc source charges the capacitors within the different modules via the charging resistors. Closing the
switch S discharge the capacitors into the parallel branches RE1, RE2 and EUT. When the EUT has a high
impedance the CWG generates a voltage rise defined by the serial inductance Ls and by the parallel
resistor RE2. The half value time is determined by Cs and the both RE1 // RE2. When the EUT has a low
impedance the generated current is determined by the elements Cs, Ls, Rs and the EUT.
The current and voltage waveforms are defined in IEC 61000-4-5 as follows:
Waveshapes and tolerances
Current into short circuit
Voltage at no load
Figure: 1.0.1-2
The waveforms are verified in open circuit (voc) and short circuit (isc). No load limitation exists, because for
different load impedance (EUT) the waveforms are within the open circuit voltage waveform and the short
circuit current waveform.
Rise time:
Ts=8 µs ± 20%
Half value time:
Tr = 20 µs ± 20%
Under swing
:
ü < 30 %
90%
50%
10%
100%
Test generator CWG
I
[A]
EUT, e.g. Motors
U
[V]
100%
50%
20 µs
8 µs
Ü<30%
50 µs
1,2 µs
Rise time:
Ts=1,2 µs ± 30%
Half value time:
Tr = 50 µs ± 20%
Over swing
:
ü < 5 %
90%
S
Summary of Contents for CDN2000-06-25
Page 18: ...CDN2000 06 25 coupling decoupling network 18 54...
Page 22: ...CDN2000 06 25 coupling decoupling network 22 54...
Page 24: ...CDN2000 06 25 coupling decoupling network 24 54...
Page 30: ...CDN2000 06 25 coupling decoupling network 30 54...
Page 36: ...CDN2000 06 25 coupling decoupling network 36 54...
Page 38: ...CDN2000 06 25 coupling decoupling network 38 54...
Page 40: ...CDN2000 06 25 coupling decoupling network 40 54...
Page 46: ...CDN2000 06 25 coupling decoupling network 46 54 V Verification by EMC PARTNER 38...
Page 47: ...CDN2000 06 25 coupling decoupling network 47 54...
Page 49: ...CDN2000 06 25 coupling decoupling network 49 54 13 3 Assembly Instruction to MC Connectors...
Page 50: ...CDN2000 06 25 coupling decoupling network 50 54...
Page 52: ...CDN2000 06 25 coupling decoupling network 52 54...
Page 53: ...CDN2000 06 25 coupling decoupling network 53 54...
Page 54: ...CDN2000 06 25 coupling decoupling network 54 54...