E-CDN2000-06-25-E-Manual
23/54
4 Panels of the CDN2000-06-25
4.1 Front panel of the CDN2000-06-25
For fusing, a combined circuit breaker is built in. The trip point lies slightly above the nominal current level
and therefore only causes a separation by a moderate continuous overload. The magnetic trips are set for a
switching point at 5 to 10 times the nominal current, so that they do not inadvertently react to the high start-
up currents of motors or to current peaks from transformers.
The SURGE coupling is clearly organised as shown in the above photograph For couplings between the
phases or phase and neutral, 18 µF is switched in as a coupling element. For couplings against the
protective earth, the coupling impedance changes to 10 Ohm with 9 µF in series
Filter and
overcurrent trip
Supply input
Supply output
Switch field coupling
impedance and path
Ls ON/OFF
Summary of Contents for CDN2000-06-25
Page 18: ...CDN2000 06 25 coupling decoupling network 18 54...
Page 22: ...CDN2000 06 25 coupling decoupling network 22 54...
Page 24: ...CDN2000 06 25 coupling decoupling network 24 54...
Page 30: ...CDN2000 06 25 coupling decoupling network 30 54...
Page 36: ...CDN2000 06 25 coupling decoupling network 36 54...
Page 38: ...CDN2000 06 25 coupling decoupling network 38 54...
Page 40: ...CDN2000 06 25 coupling decoupling network 40 54...
Page 46: ...CDN2000 06 25 coupling decoupling network 46 54 V Verification by EMC PARTNER 38...
Page 47: ...CDN2000 06 25 coupling decoupling network 47 54...
Page 49: ...CDN2000 06 25 coupling decoupling network 49 54 13 3 Assembly Instruction to MC Connectors...
Page 50: ...CDN2000 06 25 coupling decoupling network 50 54...
Page 52: ...CDN2000 06 25 coupling decoupling network 52 54...
Page 53: ...CDN2000 06 25 coupling decoupling network 53 54...
Page 54: ...CDN2000 06 25 coupling decoupling network 54 54...