E-CDN2000-06-25-E-Manual
33/54
7 Maintenance and Servicing
7.1 Maintenance
No further maintenance is necessary on the CDN2000-06-25.
7.2 Verification of the CDN2000-06-25 by the user
A simple verification whether high voltage pulses occur at the tester outputs can be carried out using an
oscilloscope of a bandwidth of 20 MHz.
7.2.1 Combination Wave Tester
Caution!
Before the verifications are started remove the connection EUT power input to public power supply.
Verification as specified in the Basic Standard 1000-4-5.
Measurement of output voltage at no load
Measurement of short circuit current with short circuit output
Check that voltage and current waveform are within the tolerances.
Calculate the source impedance from the peak voltage divided by the peak current.
1. Setting „Main Menu“
V = 2000 V; Repetition 10s
2. Measuring point:
SURGE U-CRO for the voltage measurement at no load
SURGE I-CRO for current measurement at short circuit (make a short circuit on the EUT power output
phase to neutral on the front panel of the CDN2000-06-25 generator)
3. Settings at the oscilloscope
Time base 5 µs,
Vertical deflection 0.5 V / Division
Definition of the wave forms and their tolerances see chapter Insulation
Summary of Contents for CDN2000-06-25
Page 18: ...CDN2000 06 25 coupling decoupling network 18 54...
Page 22: ...CDN2000 06 25 coupling decoupling network 22 54...
Page 24: ...CDN2000 06 25 coupling decoupling network 24 54...
Page 30: ...CDN2000 06 25 coupling decoupling network 30 54...
Page 36: ...CDN2000 06 25 coupling decoupling network 36 54...
Page 38: ...CDN2000 06 25 coupling decoupling network 38 54...
Page 40: ...CDN2000 06 25 coupling decoupling network 40 54...
Page 46: ...CDN2000 06 25 coupling decoupling network 46 54 V Verification by EMC PARTNER 38...
Page 47: ...CDN2000 06 25 coupling decoupling network 47 54...
Page 49: ...CDN2000 06 25 coupling decoupling network 49 54 13 3 Assembly Instruction to MC Connectors...
Page 50: ...CDN2000 06 25 coupling decoupling network 50 54...
Page 52: ...CDN2000 06 25 coupling decoupling network 52 54...
Page 53: ...CDN2000 06 25 coupling decoupling network 53 54...
Page 54: ...CDN2000 06 25 coupling decoupling network 54 54...