Problem/Possible Cause
*Auto-
Reset
Function
Action
Error Codes: 605
hex
(1541
dec
); 606
hex
(1542
dec
); 607
hex
(1543
dec
); 608
hex
(1544
dec
); 609
hex
(1545
dec
); 60A
hex
(1546
dec
);
60B
hex
(1547
dec
)
—
an internal CAN communication failure was detected on the flow measurement board
On the internal CAN data communica-
tion of the flow measurement board(s)
to the BioProTT
TM
FlowMCP main PCB, a
failure was detected. The internal data
communication has failed.
Electromagnetic disturbances could
lead to this failure situation.
If auto-reset function is not set:
1.
Check if there are any electromagnetic disturbances that
might influence the system. Also check the line grounding.
2.
Power off the device (disconnect the power supply) or
perform a reset of the flow channel.
3.
Power on the device again (reconnect the power supply).
→
If the error is no longer present, the device can be used like
normal.
→
If the problem persists, return the device for servicing.
If the auto-reset function is set:
The BioProTT
TM
FlowMCP resets the board to clear the error.
→
If the error is no longer present, then the device can be used
like normal.
→
If the problem persists, return the device for servicing.
Error Codes: 701
hex
(1793
dec
); 702
hex
(1794
dec
)
—
FPGA test on the flow measurement board failed during start-up
During the start-up or after a flow
channel reset, a failure on the flow
measurement board FPGA was
detected.
If auto-reset function is not set:
1.
Power off the device (disconnect the power supply) or
perform a reset of the flow channel.
2.
Power on the device again (reconnect the power supply).
→
If the error is no longer present, the device can be used like
normal.
→
If the problem persists, return the device for servicing.
If the auto-reset function is set:
The BioProTT
TM
FlowMCP resets the board to clear the error.
→
If the error is no longer present, then the device can be used
like normal.
→
If the problem persists, return the device for servicing.