EN
No.
Description
10
Possibility to set the Auto-Reset Function.
If this is set, the BioProTT
TM
FlowMCP automatically resets the flow measurement board in
an attempt to reset the error/warning when the error types/warnings affected by this are
detected. The errors/warnings for which this is relevant are indicated by a
"
" in the table in
chapter 10 "Troubleshooting").
During an automatic reset of the board, the measured flow value and/or
the RSS value of the respective flow channels may go down to zero for a few
seconds.
For more information, refer to chapter 10 "Troubleshooting".
11
Possibility to set a calibration factor within the range of 0.5 to 1.5 for each channel.
The set calibration factor is set for the flow channel. After a power cycle or connecting another
sensor, the factor remains.
For more information regarding the calibration factor as such and how to determine it, please
contact em-tec GmbH and ask for our Application Note about this subject.
12
*
Store button
Clicking this button saves all performed changes on the set-up page.
•
Please note down the changed IP address, the changed IP address mask,
the port and/or the changed password before clicking the store button.
•
Once the store button was pressed, the device can only be connected by
using the new values.
•
If the set values are unknown, please contact the service department at
em-tec GmbH.
13
Possibility to reset all performed changes on the set-up page.
Please note:
This is only possible as long as the store button was not clicked.
14
Possibility to reset a flow channel.
This can be used when e.g. an error on a flow channel is present and if this error should be
reset.
Before clicking reset, check the error according to the information given in
chapter 10 "Troubleshooting".
Table 10: Description of the set-up page of the BioProTT
TM
FlowMCP (1-4 channels)
BioProTT™ FlowMCP
Copyright em-tec GmbH
Page 37