5.8 Seals
1. Setting the parameters
To change values subject to calibration regulations (e.g. cp value), the adhesive labels on the
calibration lock in the device must be opened up and the button pressed (status "P" flashes in
the display).
2. Closing and securing the calibration lock
Once all values subject to calibration regulations have been changed, the calibration lock is
closed by pressing the button (status "P" goes out) and the access opening is sealed with an
adhesive label.
3. Securing the circuit board
The circuit board is provided with a plastic cover to protect it from tampering. The fixing
screws of this cover must be provided with an adhesive label.
4. Securing the inputs and outputs
When used in applications subject to official calibration, the terminals relevant to official
regulations (e.g. counting inputs) must be secured against tampering by calibration cover-
ing caps. Sealing is provided by an adhesive label on the screw of the covering cap.
See Chap. 5.8.1 for seal layout.
5. Securing the housing (option)
The two screws on the top of the front are designed as sealing screws.
If necessary customer can protect the housing against opening by sealing wire and a sui t-
able customer seal.
Summary of Contents for EK220
Page 2: ...Volume Conversion Device EK220 2 Elster GmbH ...
Page 110: ...Volume Conversion Device EK220 110 Elster GmbH A 2 Approval for Ex Zone ...
Page 111: ...Volume Conversion Device EK220 Elster GmbH 111 ...
Page 112: ...Volume Conversion Device EK220 112 Elster GmbH ...
Page 113: ...Volume Conversion Device EK220 Elster GmbH 113 ...
Page 114: ...Volume Conversion Device EK220 114 Elster GmbH ...