Operating instructions ekamove Rev 3 Date: June 2017 37
20 Malfunctions/Error messages
If an error occurs, an acoustic signal will sound for max. 5 minutes.
The alarm can be shut off by pressing the “STOP” key. The error
will be displayed by means of an error code.
The error code can be deleted by switching the system off and on
again.
If the power supply is interrupted during operation, the system also
goes into error mode.
If the apparatus cannot be switched on, please check the power
supply.
However, if the power is connected and the apparatus can still not
be switched on, contact the authorised ekamed specialist or notify
this problem directly to the ekamed GmbH & Co. KG company.
If the system is only pumping on one side, please check that the
supply lines are properly located and functioning. The supply lines
must not be pinched or crushed. If this is the case, remedy this
problem and start the system again. If the problem persists after
restarting, restore the system to the state in which it was delivered.
To do this, switch the system off (main switch on the bottom), then
hold down the “Start” and “Plus” keys and switch the system on.
The “-INIT-“ message appears in the display. The system is now
back in the state in which it was delivered.