Operating instructions ekamove Rev 3 Date: June 2017 18
8 Installation
8.1 Installing the air chamber system
The air chamber system is installed between the bedframe and the
soft orthopedic mattress.
When doing this, attention must be paid to the correct positioning
and attitude of the two air chambers. The system has a top side,
which can be recognised by the
“wings” of the individual air
chambers. These must face the mattress. For systems with
sensors, the top can be recognised by the sensors which are
positioned under the protective cover.
The height of the assembly should
be arranged so that the air
chamber system starts about 5 cm
above the shoulders of the person
receiving care. Please be sure to
check first how tall the person lying
in the bed is!
When the installation height has
been determined, the air chamber
system is fixed to the slatted frame
using the two fixing straps provided
for this.
Fixing straps
Air chamber system