Operating instructions ekamove Rev 3 Date: June 2017 24
9.3 Calibrating the sensors (optional)
If your system is equipped with position sensors, calibration of the
sensors is necessary after the basic parameters have been
entered. This happens automatically when you switch on, after
about 5 minutes.
For any recalibration which may be necessary, the bed is to be put
in a horizontal position, where both head and foor sections must
also be put into the horizontal position. Check that the sensors are
in the correct position (see installation of air chamber system)
By holding down the “Angle” key, the current angle of the sensors
is shown in the display.
Keeping the “Angle” key depressed,
activate the “Prog” key. “Cal.S” now appears in the display.
The sensors are now calibrated to the actual local situation.
Important!
Calibration is always necessary if the bed is changed or the
installation position of the system is altered. It is possible that
calibration may need to be carried out if the person lying on the
bed is repositioned.