Chapter6Troubleshooting
6-4. Calibration sensor and SelfCalibration problems
Problem
Possible cause and remedy
1. The calibration sensor comes out auto-
matically when the main power switch is
switched on.
• Thisisnotamalfunction.Thecalibrationsensor
comesoutautomaticallywhenthepowerisswitched
ontochecktheoperationofthecalibrationsensor.
2. The calibration sensor does not come out/
stays out.
• Turnoffthemainpower,andthenturnitonagaina
fewminuteslater.
• Ifaproblemstillremainsafterapplyingthesuggested
remedies,contactyourlocaldealer.
3. Cannot execute SelfCalibration
• Checkwhetherthedateandtimearesetcorrectly.
• Checkwhetherthemodeisset.
• Checkwhetherthecalibrationtargetsareset
correctly.
4. SelfCalibration failure
• RefertotheSelfCalibrationerrorcodetable.Ifan
errorcodethatdoesnotappearintheerrorcode
tableisdisplayed,contactyourlocaldealer.
Error Code Table
Error Code
Details
01**
• ErrorsthatoccurredduringCAL1SelfCalibration
02**
• ErrorsthatoccurredduringCAL2SelfCalibration
03**
• ErrorsthatoccurredduringCAL3SelfCalibration
09**
• ErrorsthatoccurredduringSelfCalibrationofothermodes
0*10
• Theproductmaximumbrightnessmaybelowerthanthetargetbrightness.
• Lowerthetargetbrightness.
0*34
• Thecalibrationsensormaynothavecomeoutduringcalibration,orlightmayhave
enteredthesensorduringmeasurement.
• Turnoffthemainpower,waitforafewminutesbeforeturningthepowerbackon,
andthenexecuteSelfCalibrationagain.
0*50
• Lightmayhaveenteredthecalibrationsensorwhilethesensorwasinitializing.
• ExecuteSelfCalibrationagain.