STP-iXA3306 Series Turbomolecular Pump
MT-82E-001-H
Page 106
5
S
E
R
IA
L
C
O
M
M
U
N
IC
A
T
IO
N
P
R
O
T
O
C
O
L
Parameter
Item
Data format
Remarks
1
The number of error
8-bits hexadecimal
coded ASCII
Up to 80 errors
*1
2 to 81
*1
Error 1
8-bits hexadecimal
coded ASCII
*2
…
…
Error 80
*1
8-bits hexadecimal
coded ASCII
*1 The maximum number of errors may differ depending upon the software version of the STP pump. It is recommended that
an application be designed as variable-length data. In Revision C, the case of the error detection is less than or equal 32
and Parameter length is 33.
*2 Value corresponding to the error message is transmitted, (refer to Table 36 and Table 37). The most recent error has the
largest parameter number. When the number of errors being detected is under the maximum number, the value of
parameter that is larger than the number of errors being detected is set to 0.
Example:
The number of error
: 02
hex
= 2 errors
Error 1
: 0D
hex
= 13 = Disturbance Xh
Error 2
: 0F
hex
= 15 = Disturbance Xb
Error 3 to 80
: 00
hex
= No error detected
Parameter
1
2
3
4
5
6
7
8
9
10
ASCII
0 2 0 D 0 F 0 0 0 0 0 0 0 0 0 0 0 0 0 0
HEX
30 32 30 44 30 46 30 30 30 30 30 30 30 30 30 30 30 30 30 30
Parameter
11
12
13
[O
m
itte
d]
68
69
70
71
72
73
ASCII
0 0 0 0 0 0
0 0 0 0 0 0 0 0 0 0 0 0
HEX
30 30 30 30 30 30
30 30 30 30 30 30 30 30 30 30 30 30
Parameter
74
75
76
77
78
79
80
81
ASCII
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
HEX
30 30 30 30 30 30 30 30 30 30 30 30 30 30 30 30