STP-iX455/iXL455 Series Instruction Manual
9-5
O
pe
ra
tio
n
9.2.5 Serial Communication Timeout Setting
If the signal to the input operation port of the STP pump is interrupted for a certain
period during acceleration or normal operation, the STP pump detects a failure and
stops.
The time setting of the failure detection is user definable. When setting the value to 0,
the function is disabled. This value will be common to all serial ports, and the factory
setting is 1 min.
The setting value can be changed via serial communication (see Section 9.4.15), the
STP-Link (optional accessory), or the display unit iDT-001 (optional accessory).
Design the user application so that the PC can communicate with the STP pump at
fixed regular intervals within the setting time, except when the function is disabled (the
value is 0).
Table 9.4 Serial Communication Time Out Setting
Default
Setting Range
Remark
Serial communication
time out setting
1 min.
0 to 500 min.
(1 min. step)
The function is disabled
when the value is set to 0.
◇
When the communication time out is disabled, the STP pump may not stop when
the serial communication does not function normally due to a breakage of the
communication cable. In this case, interrupt the power supply for 2 seconds or
more to stop the STP pump by power failure detection. Supply the power to the
STP pump immediately after power failure detection.
Summary of Contents for STP-iX455
Page 20: ...STP iX455 iXL455 Series Instruction Manual Installation This page intentionally blank ...
Page 56: ...STP iX455 iXL455 Series Instruction Manual Operation This page intentionally blank ...
Page 64: ...STP iX455 iXL455 Series Instruction Manual Operation This page intentionally blank ...
Page 114: ...STP iX455 iXL455 Series Instruction Manual Operation This page intentionally blank ...
Page 120: ...STP iX455 iXL455 Series Instruction Manual Troubleshooting This page intentionally blank ...
Page 124: ...STP iX455 iXL455 Series Instruction Manual Troubleshooting This page intentionally blank ...
Page 138: ...STP iX455 iXL455 Series Instruction Manual This page intentionally blank ...
Page 146: ...This page intentionally blank ...