STP-iX455/iXL455 Series Instruction Manual
9-42
O
pe
ra
tio
n
Example:
1. Remote operation mode
: 01
hex
= I/O REMOTE
3. Vent valve function
: FF
hex
= DISABLE
4. Vent valve starting speed
: 32
hex
= 50 %
5. Vent valve running time
: 003C
hex
= 60 sec.
7. Damage Limit Warning
: 00
hex
= ENABLE
8. Pump Runtime Over Warning
: FF
hex
= DISABLE
9. Pump Runtime Over Warning Hours
: 000186A0
hex
=100,000 (x 100 hours)
10. Imbalance Warning
: 00
hex
= ENABLE
12. Serial communication time out
: 003C
hex
= 60 sec.
Parameter
1
2
3
4
5
6
7
ASCII
0 1 F F F F F F 3 2 0 0 3 C F F F F 0 0
HEX
30 31 46 46 46 46 46 46 33 32 30 30 33 43 46 46 46 46 30 30
Parameter
8
9
10
ASCII
F F 0 0 0 1 8 6 A 0 0 0
HEX
46 46 30 30 30 31 38 36 41 30 30 30
Parameter
11
12
13
ASCII
F F F F F F F F F F 0 0 3 C F F F F F F
HEX
46 46 46 46 46 46 46 46 46 46 30 30 33 43 46 46 46 46 46 46
Parameter
13
ASCII
F F F F F F F F F F F F F F F F
HEX
46 46 46 46 46 46 46 46 46 46 46 46 46 46 46 46
Summary of Contents for STP-iX455
Page 20: ...STP iX455 iXL455 Series Instruction Manual Installation This page intentionally blank ...
Page 56: ...STP iX455 iXL455 Series Instruction Manual Operation This page intentionally blank ...
Page 64: ...STP iX455 iXL455 Series Instruction Manual Operation This page intentionally blank ...
Page 114: ...STP iX455 iXL455 Series Instruction Manual Operation This page intentionally blank ...
Page 120: ...STP iX455 iXL455 Series Instruction Manual Troubleshooting This page intentionally blank ...
Page 124: ...STP iX455 iXL455 Series Instruction Manual Troubleshooting This page intentionally blank ...
Page 138: ...STP iX455 iXL455 Series Instruction Manual This page intentionally blank ...
Page 146: ...This page intentionally blank ...