STP-iX455/iXL455 Series Instruction Manual
7-4
O
pe
ra
tio
n
+12V
+12V
+12V
+12V
+12V
+12V
2. 2k
Ω
2. 2k
Ω
2. 2k
Ω
2. 2k
Ω
2. 2k
Ω
2. 2k
Ω
+12V
+12V
2. 2k
Ω
X2 : Remote Input Signal
In case of Start 1)
COM
START IN
STOP IN
RESET IN
INHIBIT IN
INHIBIT ENABLE IN
PORT SELECT IN
VALVE ENABLE IN
Reset signal
Rotation inhibit signal
Input port select signal
Rotation inhibit enable signal
Vent valve enable signal
(1)
(24)
(2)
(14)
(15)
(3)
(13)
(25)
(12)
In case of Start 2)
Figure 7.2 X2 REMOTE Input Signal Pins
◇
COM pins (1, 24 pin) are isolated from the frame ground.
◇
The input current of remote input signal is approx. 5 mA. Make sure the minimum
applicable load of the relay contact when the remote operation with the relay.
◇
It is recommended to use a remote cable with shield type, and connect both
terminals to the ground.
Summary of Contents for STP-iX455
Page 20: ...STP iX455 iXL455 Series Instruction Manual Installation This page intentionally blank ...
Page 56: ...STP iX455 iXL455 Series Instruction Manual Operation This page intentionally blank ...
Page 64: ...STP iX455 iXL455 Series Instruction Manual Operation This page intentionally blank ...
Page 114: ...STP iX455 iXL455 Series Instruction Manual Operation This page intentionally blank ...
Page 120: ...STP iX455 iXL455 Series Instruction Manual Troubleshooting This page intentionally blank ...
Page 124: ...STP iX455 iXL455 Series Instruction Manual Troubleshooting This page intentionally blank ...
Page 138: ...STP iX455 iXL455 Series Instruction Manual This page intentionally blank ...
Page 146: ...This page intentionally blank ...