Note:
The address can be a decimal number from 1 to 98. The address number 0 is used to
disable multi
‑
drop mode. The address number 99 is reserved as a wild card and is used in
the query set up detailed later.
The reply will be as follows:
Reply
*
S
8
5
0
sp
r
cr
The multi
‑
drop address is stored in the pump.
To check if the pump has a multi-drop address, send the command that follows:
Command
?
S
8
5
0
cr
If the reply is as follows, the pump has multi
‑
drop mode disabled:
Reply
=
S
8
5
0
sp
0
cr
If the pump has a multi-drop address, there will be no reply. Communicate with the
pump in multi-drop message protocol. Refer to
on page 56 for
more information about multi-drop mode and multi-drop message protocol.
To check the multi-drop address of the pump, send the query as follows (using wild card
address 99 which means 'any' node):
Command
#
9
9
:
9
9
?
S
8
5
0
cr
The reply will be as follows, where dd shows the multi-drop address of the pump:
Reply
#
9
9
:
9
9
=
S
8
5
0
sp d
d
cr
Multi-drop mode can be disabled by assigning the pump an address 0. To do this, send
the command that follows (where dd shows the multi-drop address of the pump. XX
shows the address of the node that is sending the command):
Command
#
d
d
:
x
x
1
S
8
5
0
sp 0
cr
The reply will be as follows:
Reply
#
x
x
:
d
d
*
S
8
5
0
sp 0
cr
When the multi-drop mode is disabled, the pump will not respond to multi-drop
commands.
5.2 Configure the pump using a TIC
The pump can be configured using the manufacturer's TIC Turbo and Instrument
Controller or TIC Turbo Controller. Refer to
Connect the logic interface to the TIC or TAG
on page 46.
The parameters of the pump that can be set using the TIC are:
▪
Power limit setting
▪
Vent options and vent type, including operating a fan from the motor controller
▪
Standby speed setting
▪
Normal speed setting
▪
Timer settings - to enable and disable the timer, setting the time
‑
out period
07/2021 - ©Edwards Limited
Page 69
B8G000880_E
B8G000880_E - Configuration
Summary of Contents for nEXT 55
Page 110: ...This page has been intentionally left blank...
Page 114: ...This page has been intentionally left blank...
Page 115: ...This page has been intentionally left blank...
Page 116: ...edwardsvacuum com...