The timer is permanently enabled during ramp
‑
up. However, it is optional to have the
timer enabled at other times. When the pump is shipped, the Timer is enabled by
default.
To disable the timer, send the serial command that follows:
Command
!
S
8
7
0
sp
0
cr
The reply will be as follows:
Reply
*
S
8
7
0
sp
r
cr
The state of the timer option is stored in the memory of the pump.
To enable the timer again, send the serial command that follows:
Command
!
S
8
7
0
sp
1
cr
The reply will be as follows:
Reply
*
S
8
7
0
sp
r
cr
To check if the timer is enabled or disabled, send the query that follows:
Command
?
S
8
7
0
cr
The reply will be as follows (where d=0 means disabled and d=1 means enabled):
Reply
=
S
8
7
0
sp
d
cr
5.1.9 Analogue signal options
The analogue output can be used to monitor one of the five different parameters.
Table 26
Analogue signal options
Option number Description of analogue output number
0
Measured pump rotational speed. This is the factory default set-
ting.
1
Measured link power
2
Measured motor temperature
3
Measured motor controller temperature
4
Measured rotor temperature
To configure the analogue output, send the command that follows (where the character
‘d’ shows the option number):
Command
!
S
8
7
1
sp
d
cr
The reply will be as follows:
Reply
*
S
8
7
1
sp
d
cr
The analogue output signal setting is now stored in the memory of the pump.
To check which analogue output signal setting is enabled, send a query as follows:
Command
?
S
8
7
1
cr