Note:
If an external electrical load is connected to the Normal output line, the normal LED can
come on.
3.12 Pump performance data nEXT85
Table 10
Pump performance data
Parameter
nEXT85D
nEXT85H
nEXT85D
nEXT85H nEXT85D nEXT85H nEXT85D nEXT85H
Mass
3.0 Kg
3.0 Kg
4.4 Kg
4.4 Kg
3.2 Kg
3.2 Kg
2.9 Kg
2.9 Kg
Main inlet
port
ISO63
ISO63
CF63
CF63
ISO100
ISO100
NW40
NW40
Exhaust
port
NW16
Vent port
1/8 inch BSP
Main inlet pumping speed:
N
2
(l s
-1
)
84
84
84
84
86
86
47
47
He (l s
-1
)
78
78
78
78
80
78
61
61
H
2
(l s
-1
)
60
54
60
54
60
54
49
49
Ar (l s
-1
)
80
80
80
80
84
84
44
44
Peak compression ratio from the backing port to the main inlet port:
N
2
> 10
11
He
8 x 10
6
2 x 10
7
8 x 10
6
2 x 10
7
8 x 10
6
8 x 10
6
8 x 10
6
2 x 10
7
H
2
2 x 10
5
5 x 10
5
2 x 10
5
5 x 10
5
2 x 10
5
5 x 10
5
2 x 10
5
5 x 10
5
Ar
> 10
11
Ultimate
pressure
(mbar)
< 5 x 10
-9
< 5 x 10
-9
< 5 x 10
-10
< 5 x 10
-10
< 5 x 10
-9
< 5 x 10
-9
< 5 x 10
-9
< 5 x 10
-9
Note:
Pumping speeds quoted are without inlet screens. Inlet screens may reduce the pumping
speed by up to 20%. Pumping speeds are measured in accordance with ISO 5302:2003.
07/2021 - ©Edwards Limited
Page 32
B8G000880_E
B8G000880_E - Technical data
Summary of Contents for nEXT 55
Page 110: ...This page has been intentionally left blank...
Page 114: ...This page has been intentionally left blank...
Page 115: ...This page has been intentionally left blank...
Page 116: ...edwardsvacuum com...