Vent options 2
!864
0
8
decimal
Hard vent when < 50%
speed if stop or fail
?864
1
Controlled vent if > 50%
speed then hard vent if <
50% speed if stop or fail
2
If stop, hard vent
If fail, hard vent when
< 50% speed
3
If stop, hard vent
If fail, controlled vent
> 50% speed then hard
vent < 50% speed
4
If fail, hard vent
If stop, hard vent when
< 50% speed
5
If fail, hard vent
If stop, controlled vent >
50% speed then hard
vent < 50% speed
6
Hard vent if stop or fail
7
Same as option 6
8
Output is permanently
energised (Fan ON)
9
Do not use
10
Output is permanently
de-energised (Fan OFF)
11
Do not use
12
Do not use
13
Do not use
14
Hard vent when < 50%
speed if stop
15
Controlled vent if > 50%
speed then hard vent if
< 50% speed if stop.
4.6.5 Multi-drop operation
With multi
‑
drop mode, a single computer system can communicate with more than one
pump. Each pump must be assigned its own individual address before it can be installed
in a multi
‑
drop system. The command to assign the multi
‑
drop address is sent in
standard nEXT message format. Refer to
Assigning a multi-drop address
on page 68.
The message protocol in the multi
‑
drop mode is different than the protocol given for the
serial messages in the single pump systems. The differences in the multi
‑
drop message
protocol are: