SECTION 5:
MAINTENANCE
4-12
Replace the Purge Port Plug
CAUTION!
Before reinserting the Purge Port Plug, ensure it is clean and free of
debris. Even small debris can cause the unit to flood.
Note:
The shoulder on the Purge Port Plug is approximately flush with the
end cap when it is inserted fully and the glass epoxy retainer is wrapped
around the .19-inch diameter stub
1.
Purge the instrument as per the following sub-section
2.
Replace the Purge Port Plug as shown in
Figure 4-19: Replace the Purge Port Plug
3.
Replace the Purge Port Retainer using the nylon screw as shown in
Figure 4-20: The Replaced Purge Port Plug Retainer