Patient Monitor User Manual ECG/ RESP Monitoring
- 95 -
Prompt
Applicable
Patient Type
Occurring Condition
Alarm Level
COUPLET
Without
pacemaker
2 consecutive PVCs
User-selectable
BIGEMINY
Without
pacemaker
Vent Bigeminy
User-selectable
TRIGEMINY
Without
pacemaker
Vent Trigeminy
User-selectable
R ON T
Without
pacemaker
A type of single PVC under the condition
that HR<100bpm, R-R interval is less
than 1/3 the average interval, followed by
a compensating pause of 1.25X the
average R-R interval (the next R wave
advances onto the previous T wave).
User-selectable
PVC
Without
pacemaker
Single PVC detected in normal
heartbeats.
User-selectable
TACHY All
patients
Adult: 5 consecutive QRS complex, RR
interval < 0.5s.
Pediatric/neonatal: 5 consecutive QRS
complex, RR interval < 0.375s.
User-selectable
BRADY All
patients
Adult: 5 consecutive QRS complex, RR
interval
≥
1.5s .
Pediatric/neonatal: 5 consecutive QRS
complex, RR interval
≥
1s.
User-selectable
MISSED
BEATS
Without
pacemaker
When HR is less than 120 bpm., no heart
beat is tested during the period 1.75 times
of the average RR interval; or
When HR is higher than 120 bpm, no
beat is tested within 1 second.
User-selectable
IRR
Without
pacemaker
IRREGULAR RHYTHM: The patient has
irregular heart rate, check patient’s
condition, electrodes, cables and leads.
User-selectable
PNC
With
pacemaker
PACE NOT CAPTURE: After the
pacemaker is paced, QRS complex can not
be detected during 300ms.
User-selectable
PNP
With
pacemaker
PACER NOT PACED: After the QRS
complex, no pace is detected during 1.75
times of RR interval.