Patient Monitor User Manual IBP Monitoring (Optional)
- 126 -
Figure 16-3 IBP SETUP Menu
The items to be set up in the menu include:
ALM
: select
ON
to enable alarm prompt during IBP alarm. Select
OFF
to disable audio
alarm and prompt the
symbol beside
IBP
numeric.
WARNING
In order to avoid endangering the patient’s life, the user should use this function
cautiously.
ALM LEV
: set the alarm level. Three levels are available:
HIGH, MED, LOW.
ALM REC
: select
ON
to enable recording during the IBP alarm or to
OFF
to disable the
alarm recording function.
AMP ADJUST
: set to adjust waveform amplitude. Two selections are available:
MANUAL,
AUTO
. Set it to
AUTO
, the pressure names of IBP become P1 and P2, and the IBP scale is
adjusted by system automatically. Set it to
MANUAL
, the pressure names of IBP can choose
one of
ART, PA, CVP, RAP, LAP, ICP, P1, P2
and the IBP scale is adjusted by the user
via
SCALE ADJUST
item.
SWEEP
: set to select the scanning speed of the IBP wave. Two selections are available:
6.25mm/s
,
12.5 mm/s
,
25 mm/s
or
50.0mm/s
.
UNIT
: set to select the pressure unit (mmHg or kPa).
FILTER
: set this item to
12.5Hz
or
40.0 Hz
.
ALM LIMIT SETUP
: used to access the sub-menu of
IBP ALM LIMIT SETUP
, in which
the user may set up the upper and lower alarm limits of systolic pressure, diastolic pressure
and mean pressure respectively for channel 1 and channel 2.
SCALE ADJUST
: used to access the sub-menu of
IBP SCALE ADJUST
, in which the user