Patient Monitor User Manual ECG/ RESP Monitoring
- 84 -
displayed to the right side of SpO
2
. As for the sound of HR or PR in
BOTH
mode, HR is given
the priority, i.e., if HR is available, the HR sound will be sent out, but if HR is not available, then
the sound will be for PR.
There are 20s for stability before ECG measuring every time.
SWEEP
Available options for
ECG SWEEP
are 6.25, 12.5, 25.0 and 50.0 mm/s.
HR CHANNEL
"
CH1
" to count the heart rate by CH 1 waveform
"
CH2
" to count the heart rate by CH 2 waveform
LEAD TYPE
Users can select either
3 LEADS
or
5 LEADS
for this item.
ECG DISPLAY:
it varies according to
LEAD TYPE
.
When
LEAD TYPE
is set to
3 LEADS, ECG DISPLAY
can be set to
NORMAL DISPLAY
, it
can display one ECG waveform on the main screen.
When
LEAD TYPE
is set to
5 LEADS,
ECG DISPLAY
can be set to
NORMAL DISPLAY
,
MULTI-LEADS DISPLAY
and
HALF-SCN MULTI-LEADS DISPLAY
. Select
NORMAL
DISPLAY
to display two ECG waveforms on the main screen; Select
MULTI-LEADS
DISPLAY
to display seven ECG waveforms which occupying the area of seven waveforms on
the main screen; Select
HALF-SCN MULTI-LEADS DISPLAY
to display seven ECG
waveforms on the screen, occupying the area of four waveforms.
NOTE:
If
3 LEADS
is selected in the
ECG SETUP
menu, only
NORMAL DISPLAY
can be
selected for
ECG DISPLAY
item in the sub-menu.
ST ANALYSE
Pick this item to access
ST ANALYSE
. Please refer to
Section 12.7 ST segment montoring
for details.
ARR ANALYSE
Pick this item to access
ARR ANALYSE
. Please refer to
Chapter 12.8
Arr. Monitoring
for
details.
OTHER SETUP
Pick this item to access
OTHER SETUP
as shown below: