Patient Monitor User Manual Default Settings
- 170 -
A3.4 RESP
RESP Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Alarm
High
Limit
30 30 100
Alarm
Low
Limit
8 8 30
Apnea Time
20s
Calculation Type
Auto
Resp Type
Ⅱ
Sweep 12.5mm/s
Amplitude 2
A3.5 SpO
2
SpO
2
Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Alarm
High
Limit
100 100 95
Alarm
Low
Limit
90 90 88
Pitch Tone
Off
Sweep 12.5mm/s
A3.6 PR
PR Settings
ADU
PED
NEO
PR Source
SpO
2
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Alarm
High
Limit
120 160 200
Alarm
Low
Limit
50 75 100
Pulse Volume
3