Patient Monitor User Manual ECG/ RESP Monitoring
- 99 -
Lead
Ⅰ
: Placing the leads on
R-L
(RA-LA) can measure the thoracic breathing.
Lead
Ⅱ
: Placing the leads on
R-F
(RA-LL) can measure the abdominal breathing.
Figure 12-18 Install RESP leads
SWEEP
: Available options for
RESP SWEEP
are 6.25, 12.5, 25.0 and 50.0 mm/s.
WAVE AMP
: The user may set up the displaying amplitude of the RESP waveform. The
selections are
0.25/0.5/1/2/3/4/5
.
HOLD TYPE
: can be set to
AUTO
or
MANUAL
. When it is set to
AUTO
mode, the
HOLD HI
and
HOLD LO
are unavailable, and the monitor can calculate the respiration rate
automatically. When it is set to
MANUAL
mode, you can adjust the broken lines in RESP
area by the
HOLD HI
and
HOLD LO
items.
HOLD HI/LO
: when the
HOLD TYPE
is
MANUAL
, you can adjust the broken lines for
higher or lower limit of the respiration rate.
DEFAULT
: pick this item to access the
RESP DEFAULT CONFIG
dialog box, in which
the user may select whether the
FACTORY DEFAULT CONFIG
or the
USER
DEFAULT CONFIG
is to be used. After selecting any of the items and exiting the dialog
box, the system will pop up the dialog box asking for the user’s confirmation.
Figure 12-19 RESP default configuration