Patient Monitor User Manual System Menu
- 29 -
CO
2
Display CO
2
module waveform.
(If there is no CO
2
waveform on the screen, we cannot choose it. But in full
screen multi-lead mode, we can choose it though we can not see it.)
OFF
No display for this channel.
RT REC TIME
: represents “
real-time recording time
”, for which two selections are
available:
CONTINUAL
and
8S
(8 seconds). “
CONTINUAL
” means once pressing the
“
Record
” button on the front panel, the recorder will continuously print out the waveform or
parameter until the “
Record
” button is pressed again.
TIMING REC TIME
: represents “
time interval between two times of timing recording
”.
10 selections are available: “
OFF, 10MIN, 20MIN, 30MIN, 40MIN, 50MIN, 1HOUR,
2HOURS, 3HOURS
and
4HOURS
”. It means that the system will trigger the recording
operation according to the selected time interval. The recording time is fixed at 8 seconds.
NOTE:
REC TIME
has the higher priority compared with
TIMING REC TIM
E.
REC RATE
:
25.0 mm/s
or
50.0 mm/s
.
CLEAR REC TASK
: this item can be used to stop recorder from printing out too many
tasks.
NOTE:
1 The recorder is an optional part.
2 If two same waveforms are selected, one of them will change to a different waveform
automatically.
4.7 Module Setup
Select
MODULE SETUP
item in
SYSTEM SETUP
menu. You can choose the parameter you
want to monitor from this menu, so that you can enhance the display efficiency, and avoid
interference from other messages.
4.8 Tracing Waveforms Selection
Select
TRACE SETUP
item in
SYSTEM SETUP
menu to call up the following menu: