Patient Monitor User Manual NIBP Monitoring
- 115 -
Technical alarms: (display in the area below the NIBP value)
Message Cause
Alarm
Level
What to do
NIBP COMM
STOP
NIBP module failure or
communication failure
High
Stop using measuring function
of NIBP module, notify
biomedical engineer or
Manufacturer’s service staff.
NIBP AIR LEAK
NIBP pump, valve, cuff or
tube has a leakage.
Low
Check and replace the leaking
parts, if required, notify
biomedical engineer or
manufacturer’s service staff.
NIBP OVER
PRESSURE
Pressure has exceeded the
specified upper safety
limit.
Low
Measure again, if failure
persists, stop using measuring
function of NIBP module and
notify biomedical engineer or
manufacturer’s service staff.
NIBP INIT
PRESSURE
HIGH
The initial pressure is too
high during measuring
Low
Measure again, if failure
persists, stop using measuring
function of NIBP module and
notify biomedical engineer or
manufacturer’s service staff.
NIBP
SECONDARY
OVER
PRESSURE
Secondary over pressure
protection
High
Notify biomedical engineer or
manufacturer’s service staff.
NIBP TIME OUT
Measuring time has
exceeded 120 seconds
(adult) or 90 seconds
(neonatal).
Low
Measure again or use other
measuring method.
NIBP Self Test
Error
When the monitor is
powered on, NIBP
module is detected to fail
in calibration.
Low
Contact your service personnel.
NIBP CUFF
TYPE ERROR
The cuff type used isn’t
consistent with the patient
type.
Low
Confirm the patient type and
change the cuff.