Patient Monitor User Manual TEMP Monitoring
- 119 -
Chapter 15 TEMP Monitoring
15.1 TEMP Monitoring
Two TEMP probes can be used simultaneously to measure two TEMP data, and get the
temperature difference.
TEMP Monitoring Setup
With a reusable TEMP probe you can plug the probe directly into the monitor.
Apply the TEMP probes securely to the patient.
Switch on the system.
WARNING
1 Verify probe cables fault detection before the beginning of monitoring phase. Unplug
the temperature probe cable of the channe1 from the socket, and then the screen will
display the error message
TEMP1 SENSOR OFF
and the audible alarm is activated.
It is the same to the other channel.
2 Take the TEMP probe and cable carefully. When they are not in use, you should coil
up the probe and cable into a loose circle. If the wire inside the cable is tensely pulled,
it may cause mechanical damage to the probe and the cable.
15.2 TEMP SETUP
Pick the
TEMP
hot key on the screen to call up the
TEMP SETUP
menu shown as below:
Figure 15-1 TEMP SETUP
ALM
: pick
ON
to enable prompt message during the TEMP alarm; pick
OFF
to disable the