Acclarix LX4 Diagnostic Ultrasound System User Manual Troubleshooting and Maintenance
- 140 -
Figure 12-2 X, Y and the Auxiliary Idler Wheels
CAUTION
Do not drop or place foreign objects inside the trackball assembly or it may affect the
trackball operation and damage the system.
NOTE:
Be sure to clean the X and Y idler wheels and the auxiliary idler wheel.
12.3.3. Needle Guide Bracket Cleaning and Sterilization
NOTE:
1. Use proper sterilization technique at all times to perform a biopsy.
2. Ensure that protective gloves are worn.
WARNING
1. The needle guide bracket kits are not disinfected or sterilized before delivery. The
operators must clean and sterilize the needle guide bracket kits before and after each
use.
2. Inspect the bracket for damage such as cracks or breakage. If damage is evident,
discontinue use of bracket and contact your Edan representative for disposal
guidance.
3. Sterilize the bracket before disposal or sending back to manufacturer for repair.
12.3.3.1. Cleaning
1.
Wear sterile protective gloves to prevent infection.
2.
Disconnect the needle guide bracket from the transducer after each use, and remove all visible
residues from the needle guide bracket using a small and soft-bristled brush or other similar
devices. Do the cleaning quickly before the needle guide bracket dries out.
3.
Soak the needle guide bracket in the cleaning solution (Ethanol 75% or Isopropanol 70%) for
at least five minutes. Use a soft-bristled brush to clean the needle guide bracket during the
soaking.
4.
Take out the needle guide bracket from the cleanser and wipe all residues with a sterile cloth.
5.
Let the bracket air dry, or dry the bracket with a sterile cloth.
6.
If the bracket is not visually clean at the end of the cleaning steps, please repeat the cleaning
steps through step 3 to step 5.
7.
Inspect the bracket to ensure that there is no damage. The bracket should be disposed of
X idler wheel
Auxiliary idler wheel
Y idler wheel