Acclarix LX4 Diagnostic Ultrasound System User Manual Safety
- 7 -
Transducer Warnings
To avoid infection, always use protective gloves when cleaning or disinfecting
Read and follow all manufacturer instructions for disinfection agents.
To avoid infection, ensure that expiration date of the disinfecting solution has not
passed.
Disinfect the transducer after each intra-cavity or intra-operative procedure. Use a
new sterile sheath for each such procedure.
Use a pyrogen-free transducer sheath for intra-operative procedures.
The system is not intended to come into contact with the central nervous system
and the central cardiovascular system.
Unplug the transducer from the system prior to cleaning or disinfecting.
Do not immerse the transducer beyond the point indicated in
Figure 6-2.
Do not allow the transducer connector to get wet.
2.2 Cautions
Excessive dust and dirt could clog internal airflow and cause overheating. Do not
use in a dusty environment.
Inspect the system regularly, at least weekly. Before use ensure there is no visible
evidence of damage to the equipment, cables, and transducers. If a component is
damaged, replace it before use.
Do not use in locations subject to vibration.
Read and understand the section
A1.2 Ultrasound Safety and the ALARA
Principle
of Advanced User Manual before using the system. Do not expose a
patient to ultrasound energy longer than clinically reasonable.
Practice ALARA principle when operating ultrasound system. Minimize the
acoustic power without compromising the image quality.
Do not use in the presence of a flammable anesthetic.
The system generates radio frequency energy, which may cause interference with
other devices in the vicinity. If interference is suspected, try re-orienting or
relocating the equipment.
The use of electrosurgical units or other devices that generate radio frequency
interference may cause image distortion or other malfunctions.
The system should only be used by a qualified physician or allied health
professional for ultrasound evaluations.
Use only Edan supplied or recommended parts and accessories.
Verify measurement results prior to entering them into a report.
Contact your local distributor or Edan Service if there is excessive noise from the
system speaker or fans.
Please read and understand cleaning instructions prior to use.