Acclarix LX4 Diagnostic Ultrasound System User Manual Specifications
- 142 -
13 Specifications
13.1 Electrical Safety Classifications
According to the type of protection
against electric shock
Class I equipment
According to the degree of protection
against electric shock
Type BF
According to the degree of protection
against harmful ingress of liquid
Whole device: IPX0
Transducer (not including the transducer connector): IPX7;
Footswitch: IP68
According to the degree of safety of
application in the presence of a
flammable gas
Equipment not suitable for use in the presence of a flammable
gas
According to the mode of operation
Continuous operation
According to the grade of EMC
CISPR 11 Group 1, Class A
Standards Compliance
EN 60601-1:2006/A1:2013
idt IEC 60601-1: 2005/A1:2012
EN 60601-1-2: 2015
idt IEC 60601-1-2: 2014
EN 60601-2-37:2008/A1:2015
idt IEC 60601-2-37:2007/A1:2015
13.2 Power Supply
Operating Voltage
100 -240 V~
Operating Frequency
50 Hz/60 Hz
AC Input Current
2.5A-1.2A
13.3 Machine Specifications
Main unit dimensions
Height: 200mm adjustable)
Width: 680mm
Depth: 940mm(+113 adjustable)
Net weight
80kg (not including power cable or transducer)
13.4 Display Specifications
Display
TFT-LCD
Diagonal Size
21.5 inch
View Angle
Right 178°, Left 178°, Up 178°, Down 178°
Pixel Number
1920x1080
Contrast Ratio
1000 (Typical)