The preferred mode of operation when an external gas source (CO
2
or N
2
) is available is in the EXPLOSION
mode. The gas gun chamber should be charged to a pressure in excess of the well pressure before the gas gun
is exposed to well pressure (follow operating instructions in section 4). This will isolate the internal parts from the
well fluids and they will have longer life and require less maintenance.
Compact Gas Gun Disassembly and Assembly Special Precautions
When the gas gun is disassembled, always REMOVE the MICROPHONE FIRST. The microphone is located at
the lower end of the gas gun. It is removed by unscrewing a 10-32 x 1-1/4 cap screw. Then remove the Housing
Cap by unscrewing the four Allen-head cap bolts.
When reassembling the gas gun, always FIRST attach the Housing Cap to the housing with the 4 bolts then
INSTALL the MICROPHONE LAST. Do not over torque bolts.
The four cap bolts, Part No. GG-210, which hold the housing cap to the main gas gun housing should be
properly tightened and periodically replaced. The proper torque for the bolts is approximately 50 inch-pounds.
The "O" ring seal between the housing cap and the housing does not require excessive tightening of the bolts to
properly seal. Do not over tighten bolts. If the bolts are excessively tightened, the bolts would be pre-tensioned to
a stress such that when internal pressure is applied, the additional force on the housing cap generated by the
internal pressure could cause the bolts to fail.