Rarefaction (Implosion) Mode
Gas is released from the well into the gas gun volume chamber to generate the initial pulse. Debris, moisture,
corrosive liquids and chemicals, and other foreign material may be imploded into the gas gun volume chamber,
which will increase maintenance requirements and may cause corrosion on the inside of the volume chamber.
•
Securely attach the Echometer Gas Gun to the Casing or Tubing Valve.
•
Close the Well Pressure Bleed Valve and Filler Bleed Chamber Valve.
•
Open the Well Valve to the Echometer Gas Gun slowly and close the casing valve or the wing valve to
the flow line.
•
Pull Trigger Ring.
•
Lift the Cocking Arm to close the internal gas valve.
•
Measure the Well Pressure using the precision pressure gauge.
•
Record Time and Well Pressure.
•
Bleed the gas chamber pressure through the Filler-Bleeder Chamber Valve by rotating the knob
clockwise until the gas gun pressure has decreased to approximately 200-psi below the well pressure
reading. Use greater or less differential pressure depending on the background noise level.
•
Connect the coaxial cable from the microphone to the INPUT of the Model H.
•
Power up Instrument by pressing left soft key.
•
Navigate to the Acoustic Record Acquisition screen following the Implosion Pulse path as described in
the following section.
•
Generate pressure pulse by pulling Trigger Ring, after the message “Waiting for Shot to be Fired” is
displayed.
•
Inspect the record and repeat the shot if the signal quality is not satisfactory.
•
Record Time and Well Pressure.
•
Close the Casing Valve or Tubing Valve between Echometer Gas Gun and the well.
•
Open the Well Pressure Bleed Valve and release the pressure.
•
Open the Casing Valve or Wing Valve to the flow line.
•
Remove the Echometer Gas Gun from the Casing or Tubing valve.