XS series operating manual
XS user manual v0.55
- 69 -
The red lamp
« Danger, HV ON »
lights and the display shows ( in example) :
Fig 8.19
IMPORTANT :
ALWAYS MAKE SURE THAT THE HIGH VOLTAGE INDICATOR (
RED LAMP
) IS NOT ON WHEN
CONNECTING OR DISCONNECTING THE SPECIMEN.
In case where the MANUAL test time mode has been selected, the unit remains in the measurement
mode permanently until the [DISCHARGE] push button is pressed.
Use the UP arrow key to increase the output voltage and the DOWN arrow key to decrease the output
voltage. The maximum voltage which can be reached is those which is displayed on the parameter
line :
VOLTAGE: x.xx KVAC
However according to the impedance of the specimen under test, the maximum reached
voltage could be lower.
If the AUTO test time has been selected, every second the value of the RISE time counts down one
unit, then the value of the HOLD time and at least the value of the FALL time. When the FALL time
reaches 0 the output voltage is cut off automatically. According to the leakage current value during the
test time in comparison with the IMAX and IMIN limits, the red LED (
FAIL
) or the green LED (
PASS
) is
illuminated.