XS series operating manual
XS user manual v0.55
- 126 -
DET(DETection) OFF / I / FI / I+DELTA / FI+DELTA / DELTA :
Hipot function only sets the mode of the breakdown.
IDE(IDElta) <NR3>
Hipot function : Delta I breakdown detection value adjustment
[expert]
RDET(RDETection) OFF/ I / I+DELTA / DELTA / FI / FI+DELTA
Hipot function: Selection of the breakdown detection mode during the high voltage rise
and fall times.
[expert]
IMRA <NR3>
Hipot function:setting of the break detection value IMAX or FIMAX during the high voltage
rise and fall times.
[expert]
IDERA <NR3>
Hipot function:setting of the break detection value of Delta I during the high voltage rise
and fall times.
[expert]
UDISP ON/OFF
Hipot function: Display control of the value of high voltage during the high voltage rise
and fall times.
[expert]
IDISP ON/OFF
Hipot function: Display control of the value of the current during the high voltage rise and
fall times.
MEAS(MEASure)
Runs the current function.
STOP
Stops the current function.
QUIT
Exits the current function.
SRQ
Similar to the Service Request feature of the IEEE488-2 bus, this code allows the
emission of the
Z
character to inform the computer about the events (end of test, format
error, interlock open,...). This command has to be sent at the program start, after the REM
command.
Device queries
MEAS?
Return the current measurement value(s) regarding the running function.
A) Following characters in the function
Insulation
: OHM 4.700E+06
Dielectric strength
: VOLT 9.900E+02 AMP 7.000E-05
Continuity
: OHM 3.210E-1 VOLT 2.810E+00 (main unit =OHM)