XS series operating manual
XS user manual v0.55
- 28 -
4 Ground continuity function
Measurement current
From 5 to 32A AC ( 40A for CXS 40 ) by step of 0.5 A AC.
Frequency = mains frequency
Accuracy :
(1% + 500mA).
Open circuit voltage
< 6 VAC or < 12 VAC.( < 9V for CXS40 )
Sinus wave.
Frequency identical to the mains.
Measurement accuracy
Display on 1500 points digital indicator with unity indications (m
).
Accuracy (in % of the reading. 1U = 1 display count 1m
) :
(2.5%+10U) in the following
measurement range :