XS series operating manual
XS user manual v0.55
- 96 -
Information on cycle timing
Minimum cycle timing in dielectric strength test ( hipot test)
Measurement conditions :
Software revision
Measurement display NO (second page of the setup menu)
Interface
PLC
Rise time
0
Fall time
0
Hold time
1
Time
AUTO
Other parameters
independent
Test time = time between change of the MES_DCH (start test) signal and change of EOT (end of
test) signal =
980 mS
Measurement parameters :
same as above but with hold time = 0
Test time =
700 mS.
In this case there is no output voltage control.
Measurement parameters :
memory number change between each measurement
Test time = basic minimum test time +
1.3 S
Minimum cycle timing in MEGOHMMETER
Measurement conditions :
Software revision
Measurement display
NO (second page of the setup menu)
Interface
PLC
Measurement time
0
Other parameters
independent
Test time = time between change of the MES_DCH (start test) signal and change of EOT (end of
test) signal =
1.58 S
Measurement parameters :
memory number change between each measurement
Test time = basic minimum test time +
1.3 S
Minimum cycle timing in GROUND CONTINUITY
Measurement conditions :
Software revision
Measurement display
NO (second page of the setup menu)
Interface
PLC
Rise time
0
Fall time
0
Hold time
1
Time
AUTO
Test time = time between change of the MES_DCH (start test) signal and change of EOT (end of test)
signal =
1.19 S
Measurement parameters :
memory number change between each measurement
Test time = basic minimum test time +
1.3 S